Binder | Model BD-S 115 | Standard-Incubators with natural convection

Download

Binder | Model BD-S 115 | Standard-Incubators with natural convection

 

Binder | Model BD-S 115 | Standard-Incubators with natural convection

 

Binder Test Cihazları Türkiye Distribütörü | Enfor

 

Temperature range: +5 °C above ambient temperature to +70 °C
APT.line™ preheating chamber technology
Natural convection
Adjustable exhaust air flap
Controller with timer function
Inner door made of tempered safety glass
1 chrome-plated rack incl. shelf supports
Class 3.1 integrated independent temperature safety device (DIN 12880) with visual alarm

 

Description BDS115-230V BDS115UL-120V
     
Article Number 9090-0022 9090-0023
Performance Data Temperature    
Temperature range +5 °C above ambient temperature to [°C] 70 70
Temperature variation at 37 °C [± K] 0.5 0.5
Temperature fluctuation at 37 °C [± K] 0.3 0.3
Recovery time after 30 seconds door open at 37 °C [min] 15 15
Electrical data    
Rated Voltage [V] 230 120
Power frequency [Hz] 50/60 60
Nominal power [kW] 0.35 0.35
Unit fuse [A] 6.3 6.3
Phase (Nominal voltage) 1~ 1~
Measures    
Interior volume [L] 118 118
Net weight of the unit (empty) [kg] 47 47
Permitted load [kg] 75 75
Load per rack [kg] 20 20
Wall clearance back [mm] 160 160
Wall clearance sidewise [mm] 100 100
Internal Dimensions    
Width [mm] 550 550
Height [mm] 550 550
Depth [mm] 390 390
Doors    
Inner doors 1 1
Unit doors 1 1
Housing dimensions not incl. fittings and connections    
Width net [mm] 764 764
Height net [mm] 735 735
Depth net [mm] 616 616
Environment-specific data    
Energy consumption at 37 °C [Wh/h] 25 25
Fixtures    
Number of shelves (std./max.) 1/5 1/5