F32-ME Refrigerated Heating Circulator

Download

F32-ME Refrigerated Heating Circulator

Technical Data;
Working temperature range (°C) :  -35 ... +200
Temperature control : PID, cascade
Temperature stability (°C) : ±0.01
Setting / display resolution : 0.01 °C
Integrated programmer : 1x10 steps
Temperature Display : VFD
Heating capacity (kW) : 2
Cooling capacity (Medium Ethanol) @20 °C : 0.45 kW 
Pump capacity flow rate (l/min) : 11-16
Pump capacity flow pressure (bar) : 0.23-0.45
Bath opening / bath depth (W x L / D cm) : 18 x 12 / 15
Pump connections : M10x1
Barbed fittings diameter (inner dia. / mm) : 8 / 10
Filling volume (liters) : 8
Refrigerant stage 1 : R134a
Filling volume refrigerant stage 1 (g) : 166
Global Warming Potential for R134a : 1430
Carbon dioxide equivalent stage 1 (t) : 0.237
External Pt100 sensor connection : integrated
Digital interfaces : RS232
Optional : Profibus
Ambient temperature : 5...40 °C
Dimensions W x L x H (cm) : 31 x 42 x 64
Weight (kg) : 37
Classification according to DIN12876-1 : Classification III (FL)
Cooling of compressor : Air