Technical Data;
Working temperature range (°C) : -50 ... +200
Temperature control : ICC
Temperature stability (°C) : ±0.01
Setting / display resolution : 0.01 °C
Integrated programmer : 1x10 steps
Temperature Display : VFD
Heating capacity (kW) : 2
Cooling capacity (Medium Ethanol) @ 20 °C : 0.9 kW
Pump capacity flow rate (l/min) : 22-26
Pump capacity flow pressure (bar) : 0.4-0.7
Pump capacity flow suction (bar) : 0.2-0.4
Bath opening / bath depth (W x L / D cm) : 18 x 12 / 15
Pump connections : M16x1
Barbed fittings diameter (inner dia. / mm) : 8 / 12
Filling volume (liters) : 8
Refrigerant stage 1 : R404A
Filling volume refrigerant stage 1 (g) : 480
Global Warming Potential for R404A : 3922
Carbon dioxide equivalent stage 1 (t) : 1.883
External Pt100 sensor connection : integrated
Digital interfaces : RS232
Optional : Profibus
Ambient temperature : 5...40 °C
Dimensions W x L x H (cm) : 42 x 49 x 72
Weight (kg) : 57
Classification according to DIN12876-1 : Classification III (FL)
Cooling of compressor : Water