Binder KMF 720 Model Constant Climate Chamber with Expanded Temperature / Humidity Range
Binder
Model KMF 115 | Constant Climate Chamber with Expanded Temperature / Humidity Range
| * Temperature range: -10 °C to +100 °C * Humidity range: 10 % RH to 98 % RH * APT.line™ preheating chamber technology * Humidity regulation with capacitive humidity sensor and vapor humidification * Inner chamber made of stainless steel * BINDER Multi Management Software APT-COM™ Basic Edition * Intuitive touchscreen controller with time-segment and real-time programming * Internal data logger, measured values can be read out in open format via USB * Unit self-test for comprehensive status analysis * Tight-sealing inner door made of safety glass (ESG) * Avoidance of glass corrosion by special TIMELESS coating * 1 stainless steel rack * Access port with silicone plug, 30 mm, left * Class 3.1 independent temperature safety device (DIN 12880) with visual and audible temperature alarm * Computer interface: Ethernet * Door heating |
| Description | KMF720-230V * | KMF720-240V * |
| Article Number | 9020-0345 | 9020-0346 |
| Performance Data Temperature | ||
| Temperature range | -10…100 | -10…100 |
| Average heating-up rate according to IEC 60068-3-5 [K/min] | 0.7 | 0.7 |
| Average cooling down time according to IEC 60068-3-5 [K/min] | 0.4 | 0.4 |
| Max. heat compensation at 25 °C [W] | 450 | 450 |
| Performance Data Climate | ||
| Temperature range [°C] | 10…90 | 10…90 |
| Temperature variation depending on setpoint [± K] | 0.1…1 | 0.1…1 |
| Temperature fluctuation depending on setpoint [± K] | 0.1…0.5 | 0.1…0.5 |
| Humidity range | 10…98 | 10…98 |
| Humidity fluctuation depending on setpoint | ≤2 ± % RH | ≤2 ± % RH |
| Dew point temperature range [°C] | 5…90 | 5…90 |
| Electrical data | ||
| Rated Voltage [V] | 200…230 | 200…240 |
| Power frequency [Hz] | 50/60 | 50/60 |
| Nominal power [kW] | 3.1 | 3.1 |
| Unit fuse [A] | 16 | 16 |
| Phase (Nominal voltage) | 1~ | 1~ |
| Measures | ||
| Interior volume [L] | 700 | 700 |
| Net weight of the unit (empty) [kg] | 306 | 306 |
| Permitted load [kg] | 150 | 150 |
| Load per rack [kg] | 45 | 45 |
| Wall clearance back [mm] | 100 | 100 |
| Wall clearance sidewise [mm] | 100 | 100 |
| Internal Dimensions | ||
| Width [mm] | 973 | 973 |
| Height [mm] | 1250 | 1250 |
| Depth [mm] | 576 | 576 |
| Doors | ||
| Inner doors | 2 | 2 |
| Unit doors | 2 | 2 |
| Housing dimensions not incl. fittings and connections | ||
| Width net [mm] | 1250 | 1250 |
| Height net [mm] | 1925 | 1925 |
| Depth net [mm] | 890 | 890 |
| Environment-specific data | ||
| Energy consumption at 85 °C and 85 % RH [Wh/h] | 900 | 900 |
| Sound-pressure level [dB(A)] | 56 | 56 |
| Fixtures | ||
| Number of shelves (std./max.) | 1/15 | 1/15 |