Binder | Model ED 400 | Drying and Heating Chambers with Natural Convection
Binder | Model ED 400 | Drying and Heating Chambers with Natural Convection
| * Temperature range: +5 °C above ambient temperature to +300 °C * APT.line™ preheating chamber technology * Natural convection * Adjustable exhaust air flap * Controller with timer functions * 2 chrome-plated racks * Class 2 independent adjustable temperature safety device (DIN 12880) with visual alarm |
| Description | ED400-400V-I * | ED400UL-208V-I * |
| Article Number | 9010-0075 | 9010-0168 |
| Performance Data Temperature | ||
| Temperature range +5 °C above ambient temperature to [°C] | 300 | 300 |
| Temperature variation at 150 °C [± K] | 3 | 3 |
| Temperature fluctuation at 150 °C [± K] | 0.5 | 0.5 |
| Heating-up time to 150 °C [min] | 71 | 71 |
| Recovery time after 30 seconds door open at 150 °C [min] | 31 | 31 |
| Air change data | ||
| Air change (approx.) at 150 °C [x/h] | 10 | 10 |
| Electrical data | ||
| Rated Voltage [V] | 400 | 208 |
| Power frequency [Hz] | 50/60 | 60 |
| Nominal power [kW] | 3.4 | 3.4 |
| Unit fuse [A] | 3 x 16 | 3 x 16 |
| Phase (Nominal voltage) | 3~ | 3~ |
| Measures | ||
| Interior volume [L] | 400 | 400 |
| Net weight of the unit (empty) [kg] | 125 | 125 |
| Permitted load [kg] | 90 | 90 |
| Load per rack [kg] | 35 | 35 |
| Wall clearance back [mm] | 160 | 160 |
| Wall clearance sidewise [mm] | 100 | 100 |
| Internal Dimensions | ||
| Width [mm] | 1000 | 1000 |
| Height [mm] | 800 | 800 |
| Depth [mm] | 520 | 520 |
| Doors | ||
| Unit doors | 2 | 2 |
| Housing dimensions not incl. fittings and connections | ||
| Width net [mm] | 1235 | 1235 |
| Height net [mm] | 1025 | 1025 |
| Depth net [mm] | 765 | 765 |
| Environment-specific data | ||
| Energy consumption at 150 °C [Wh/h] | 672 | 672 |
| Fixtures | ||
| Number of shelves (std./max.) | 2/10 | 2/10 |