Binder | Model FD 23 | Drying and Heating Chambers with Forced Convection
Binder | Model FD 23 | Drying and Heating Chambers with Forced Convection
| * Temperature range: +5 °C above ambient temperature to +300 °C * APT.line™ preheating chamber technology * Forced convection * Adjustable exhaust air flap * Controller with timer functions * 2 chrome-plated racks * Class 2 independent adjustable temperature safety device (DIN 12880) with visual alarm |
| Description | FD023-230V * | FD023UL-120V * |
| Article Number | 9010-0194 | 9010-0196 |
| Performance Data Temperature | ||
| Temperature range +5 °C above ambient temperature to [°C] | 300 | 300 |
| Temperature variation at 150 °C [± K] | 2.5 | 2.5 |
| Temperature fluctuation at 150 °C [± K] | 0.3 | 0.3 |
| Heating-up time to 150 °C [min] | 25 | 25 |
| Recovery time after 30 seconds door open at 150 °C [min] | 6 | 6 |
| Air change data | ||
| Air change (approx.) at 150 °C [x/h] | 64 | 64 |
| Electrical data | ||
| Rated Voltage [V] | 230 | 120 |
| Power frequency [Hz] | 50/60 | 60 |
| Nominal power [kW] | 0.8 | 0.8 |
| Unit fuse [A] | 10 | 12.5 |
| Phase (Nominal voltage) | 1~ | 1~ |
| Measures | ||
| Interior volume [L] | 20 | 20 |
| Net weight of the unit (empty) [kg] | 27 | 27 |
| Permitted load [kg] | 25 | 25 |
| Load per rack [kg] | 12 | 12 |
| Wall clearance back [mm] | 160 | 160 |
| Wall clearance sidewise [mm] | 100 | 100 |
| Internal Dimensions | ||
| Width [mm] | 222 | 222 |
| Height [mm] | 330 | 330 |
| Depth [mm] | 300 | 300 |
| Doors | ||
| Unit doors | 1 | 1 |
| Housing dimensions not incl. fittings and connections | ||
| Width net [mm] | 435 | 435 |
| Height net [mm] | 495 | 495 |
| Depth net [mm] | 520 | 520 |
| Environment-specific data | ||
| Energy consumption at 150 °C [Wh/h] | 300 | 300 |
| Fixtures | ||
| Number of shelves (std./max.) | 2/4 | 2/4 |