Binder | Model FP 720 | Drying and Heating Chambers with Forced Convection and Program Functions
Binder | Model FP 115 | Drying and Heating Chambers with Forced Convection and Program Functions
| * Temperature range: +5 °C above ambient temperature to +300 °C * APT.line™ preheating chamber technology * Adjustable fan speed * Adjustable exhaust air flap * Controller with time-segment and real-time programming * 2 chrome-plated racks * Class 2 independent adjustable temperature safety device (DIN 12880) with visual alarm * Computer interface: RS 422 |
| Description | FP720-400V * | FP720UL-208V * |
| Article Number | 9010-0267 | 9010-0268 |
| Performance Data Temperature | ||
| Temperature range +5 °C above ambient temperature to [°C] | 300 | 300 |
| Temperature variation at 150 °C [± K] | 2 | 2 |
| Temperature fluctuation at 150 °C [± K] | 0.3 | 0.3 |
| Heating-up time to 150 °C [min] | 39 | 39 |
| Recovery time after 30 seconds door open at 150 °C [min] | 20 | 20 |
| Air change data | ||
| Air change (approx.) at 150 °C [x/h] | 12 | 12 |
| Electrical data | ||
| Rated Voltage [V] | 400 | 208 |
| Power frequency [Hz] | 50/60 | 60 |
| Nominal power [kW] | 5 | 5 |
| Unit fuse [A] | 3 x 16 | 3 x 16 |
| Phase (Nominal voltage) | 3~ | 3~ |
| Measures | ||
| Interior volume [L] | 720 | 720 |
| Net weight of the unit (empty) [kg] | 194 | 194 |
| Permitted load [kg] | 120 | 120 |
| Load per rack [kg] | 45 | 45 |
| Wall clearance back [mm] | 160 | 160 |
| Wall clearance sidewise [mm] | 100 | 100 |
| Internal Dimensions | ||
| Width [mm] | 1000 | 1000 |
| Height [mm] | 1200 | 1200 |
| Depth [mm] | 610 | 610 |
| Housing dimensions not incl. fittings and connections | ||
| Width net [mm] | 1235 | 1235 |
| Height net [mm] | 1530 | 1530 |
| Depth net [mm] | 865 | 865 |
| Environment-specific data | ||
| Energy consumption at 150 °C [Wh/h] | 1320 | 1320 |
| Fixtures | ||
| Number of shelves (std./max.) | 2/16 | 2/16 |